◤OVERVIEW
The HOBO UX100-014M Single Channel Thermocouple data logger records temperature in indoor environments with thermocouple sensors (sold separately). Designed for a broad range of monitoring applications – from industrial process monitoring to concrete maturity testing – the logger provides exceptional flexibility by accepting J, K, T, E, R, S, B or N type thermocouple probes.
Onset also offers a four-channel thermocouple data logger, the HOBO UX120-014M.
- High-accuracy J, K, T, E, R, S, B, N thermocouple readings
- Easy-to view LCD display
- Large memory capacity
- Visual high & low alarm thresholds
- New Burst and Statistics logging modes
- Record ambient temperature with internal thermistor
In what environment does this data logger operate?
This data logger operates in an indoor environment.
What measurements does this data logger support?
The UX100-014M data logger supports the following measurement:Temperature
Thermocouple
- Type:J
Range:-210° to 760°C (-346° to 1,400°F)
Accuracy:±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution:0.03°C (0.06°F) - Type:K
Range:-260° to 1,370°C (-436° to 2,498°F)
Accuracy:±0.7°C (±1.26°F) ± thermocouple probe accuracy
Resolution:0.04°C (0.07°F) - Type:T
Range:-260° to 400°C (-436° to 752°F)
Accuracy:±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution:0.02°C (0.03°F) - Type:E
Range:-260° to 950°C (-436° to 1,742°F)
Accuracy:±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution:0.03°C at (0.05°F) - Type:R
Range:-50° to 1,550°C (-58° to 2,822°F)
Accuracy:±2.2°C (±3.96°F) ± thermocouple probe accuracy
Resolution:0.08°C (0.15°F) - Type:S
Range:-50° to 1,720°C (-58° to 3,128°F)
Accuracy:±2.2°C (±3.96°F) ± thermocouple probe accuracy
Resolution:0.08°C (0.15°F) - Type:B
Range:550° to 1,820°C (1,022° to 3,308°F)
Accuracy:±2.5°C (±4.5°F) ± thermocouple probe accuracy
Resolution:0.1°C (0.18F) - Type:N
Range:-260° to 1,300°C (-436° to 2,372°F)
Accuracy:±1.0°C (±1.8°F) ± thermocouple probe accuracy
Resolution:0.06°C (0.11°F)
- Range:Logging:-20° to 70°C (-4° to 158°F)
- Accuracy:±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
- Resolution:0.024°C at 25°C (0.04°F at 77°F); see Plot A
- Drift:<0.1°C (0.18°F) per year
- Logger Operating Range
。Logging:-20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); logger accuracy from 0° to 50°C (32° to 122°F)
。Launch/readout:0° to 50°C (32° to 122°F) per USB specification - Resolution:22-bit
- Logging rate:1 second to 18 hours, 12 minutes, 15 seconds
- Logging modes:Normal, Burst or Statistics
- Memory modes:Wrap when Full or Stop when Full
- Start modes:Immediately, Push Button, Date & Time or Next Interval
- Stop modes:When Memory Full, Push Button, or Date & Time
- Restart mode:Push Button
- Time Accuracy - UX120-014MTime accuracy:±1 minute per month at 25°C (77°F), see Plot B
- Battery life:1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater
- Battery type:One 3V CR2032 lithium battery
- Memory:512 KB (208,076 measurements, maximum)
- Download type:USB 2.0 interface
- Full memory download time:1 minute, 30 seconds
- LCD:LCD is visible from:0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
- Size:3.66 x 5.94 x 1.52 cm (1.44 x 2.34 x 0.6 in.)
- Weight:30 g (1.06 oz)
- Environmental rating:IP50
– The CE Marking identifies this product as complying with all relevant directives in the European Union (EU)